Clamp Sensor

Clamp Sensor


Clamps on Probe 3275 Sensors for oscilloscope / memory hicorders, DC to 100 mhz wide-band for observing waveforms. Even with the wide DC to 2 mhz frequency band, current waveforms up to 500 A can be directly observed in applications such as power device and power-on in-rush current monitoring. An optional power supply is available for use with oscilloscopes that do not provide probe power.

  • Frequency band width: DC to 2 MHz (-3 dB)
  • Rise time: =<175 ns
  • Noise level: 25 mA rms max. (band width limited to 20 MHz)
  • Continuous allowable input: 500 A rms (requires derating at frequency)
  • Max. allowable peak input: 700 A peak (non continuous)
  • Output rate: 0.01 V/A (use with a device having a 1 M-ohm input resistance or higher)
  • Max. rated voltage to earth: 600 V CAT II, 300 V CAT III (insulated wire)
  • Core jaw dia: Ø20 mm(0.79 in)
  • Power supply: ±12 V ±0.5 V, 7.2 VA max. (at Max. current measurement)
  • Power plug: FFA.0S.304.CLAC42Z / LEMO inc.
  • Supplied accessories: Carrying case × 1

  • Connects directly to oscilloscope BNC input terminal
  • Connects directly to the MEMORY HiCORDER to provide waveform monitoring
  • DC to 2MHz (-3dB)
  • 500Arms maximum
  • 0.01V/A output
  • ?20mm (0.79in) dia.