Calibration Of Dead Pressure Gauge

Calibration Of Dead Pressure Gauge


Gujarat Instrumentation Institute offers Calibration of Dead Pressure Gauge. Recommended calibration reference temp. is 27 ± 5 'C and relative humidity not more than 85% RH. Calibration area should be adequately free from dust, shock and vibration. Accuracy ratio of master gauge and gauge under test should be 3, 5 or 10 times higher depending upon availability and required calibration standard. Care should be taken for oxygen service pressure gauges (Use no oil). Pressure gauge is calibrated with using comparison test pump and master pressure gauge. (Digital Precision Pressure Calibrator. Recommended calibration reference temp. is 27 ± 5 'C and relative humidity not more than 85% RH. Calibration area should be adequately free from dust, shock and vibration. Accuracy ratio of master gauge and gauge under test should be 3, 5 or 10 times higher depending upon availability and required calibration standard. Care should be taken for oxygen service pressure gauges (Use no oil). Pressure gauge is calibrated with using comparison test pump and master pressure gauge. (Digital Precision Pressure Calibrator.